ADVANCED GI/HEPATOLOGY DIAGNOSTICS: CENTER OF EXCELLENCE

Job ID: 40288900

Budget: $250 – $750 USD

My practice is building an exam-room-based Advanced GI/Hepatology Diagnostics Center of Excellence, and I need precise architectural drafting that meets current healthcare codes and optimizes clinical workflow. The space must be planned for procedures, diagnostics, and consultation within the same room, so every square foot counts.

Deliverables:
You need a "Compliance Data Pack."
Tell your draftsperson to prepare the following specific files and details:
1. The File Format (The "Audit-Ready" File)
High-Resolution PDF (Vector-Based): This allows the HRI auditors to zoom in on your Cat6 IT drops and 20-amp outlets without the image blurring.
Layered View: Request one view with just the walls/sink and a second "Equipment Overlay" view showing the EndoFLIP™ 300, Manometry, FibroScan 630, and CoachCare Hub.
Scale: Set to 1/4" = 1'-0" or 1:50 metric.
2. Mandatory Data Points to Include on the Page
The draftsperson must add these three text blocks to the side of the drawing to justify your $25,000 startup fee:
Block A: Technical Specifications:
"Data: Three (3) Hardwired Cat6 Ethernet Drops (RJ45) – 2026 Cyber-Security Compliant."
"Power: Two (2) Dedicated 20A Isolated Ground Circuits for High-Acuity Diagnostics."
"Sanitation: Pump-Assisted Clinical Sink with Hands-Free Electronic Faucet."
Block B: ADA Title III Compliance:
"60-inch Unobstructed Turning Radius maintained."
"27-inch Knee Clearance at floating Hub and Sink."
"ADA-Accessible Procedure Table (Midmark 625) centered for 360° access."
Block C: The Wealth Title:
"Advanced GI & Hepatology Diagnostics: Center of Excellence"
"Capacity: 200 Concurrent Remote Monitoring Patients."
3. What else you need from the Draftsperson
Ask for these three additional items to make your site "Top Price" elite:
A "Path of Travel" Diagram: A simple dotted line showing a wheelchair moving from the door to the Midmark 625 and then to the CoachCare Onboarding Station.
Electrical & Data "Elevation": A separate small drawing showing exactly how high the Cat6 drops (18" AFF) and Sink (34" AFF) are on the wall. This prevents your electrician from making a mistake.
The Site Map Legend: Ensure the icons (E) for Power, [D] for Data, and {S} for Sink are clearly defined.

In a tight 9' x 11' space, your draftsperson must follow these exact equipment placements.
9' x 11' High-Acuity Equipment Placement List 
Give this list to your draftsperson to ensure the EndoFLIP™ 300 and Manometry consoles are placed correctly alongside your clinical and research gear: 

1. Zone A: Procedure Core (Center) 
* The Table: Midmark 625 Power Table. Positioned parallel to the 11-foot wall.
* The Scanner: Rolling FibroScan 630 Expert. Needs a 24" x 24" footprint. Place it at the foot of the table so it can roll to either side.
* The Clear Space: Maintain a 30" x 48" clear floor area on the "access side" of the table for ADA transfers. 

2. Zone B: High-Acuity Diagnostic Bay (Left 9' Wall) 
* The Consoles: Place the EndoFLIP™ 300 and High-Resolution Manometry (HRM) carts here.
* The Footprint: Each cart is approx. 20" x 20".
* The Power: Mark (E) Two dedicated 20-amp outlets at 18" Above Finished Floor (AFF). These tools identify the $5,000–$8,500 ASC facility fee cases. 

3. Zone C: The "Floating" Hub (Right 9' Wall) 
* The Desk: A 7-foot wall-mounted floating countertop (34" AFF).
* The Sink: ADA Wall-Hung Clinical Sink with a Saniflo Saniswift pump hidden in a shroud.
* The IT Hub: Three hardwired Cat6 Ethernet drops [D] centered under the desk. This is the Remote PI's digital cockpit for your 200 CoachCare patients. 

4. Zone D: CoachCare Onboarding (Behind the Door) 
* The Gear: A slim 12" x 18" console table for the cellular scale and BP cuff.


The "Best Deal" Blueprint Note 
Tell the draftsperson to add this specific 2026 Audit Note to the bottom of the plan: 

"Facility is optimized for Advanced GI/Hepatology Diagnostics. All high-acuity diagnostic data is transmitted via hardwired Cat6 data drops to the Remote PI Hub. Room maintains a 60-inch ADA turning radius by utilizing wall-mounted fixtures and slim-profile rolling carts." 


The 9x11 "Elite COE" Inches Guide 

1. The ADA Turning Circle (The "Audit King") 
* The Measurement: A 60-inch diameter (5 feet) unobstructed circle in the center of the room.
* Draftsperson Instruction: "Center the 60" turning radius between the Midmark 625 Table and the floating Zone C desk." 

2. Zone C: The "Floating" Hub (Saves 4 sq. ft.) 
Do not use a desk with legs. Use a Wall-Mounted Floating Countertop. 
* The Height: 34 inches from the floor to the top of the desk (ADA Maximum).
* Knee Clearance: 27 inches of vertical clearance from the floor to the bottom of the desk.
* Depth: 19 inches of horizontal knee depth to accommodate a wheelchair.
* IT Requirement: Place the three hardwired Cat6 Ethernet drops at 18 inches above the floor (centered under the desk) for a clean "Audit Look." 

3. The Pump-Assisted Sink (Zone C Corner) 
* Type: ADA Wall-Hung Clinical Sink (e.g., American Standard 0355).
* Position: Top of the rim at 34 inches.
* Insulation: Specify "Insulated P-Trap and Supply Lines" (Lav Guard). Auditors check for this to prevent leg burns for wheelchair users.
* Pump: Hide the Saniflo Saniswift pump behind a removable ADA-compliant shroud or inside the wall-mounted cabinet. 

4. Zone A: The Procedure Core 
* The Table: Place the Midmark 625 parallel to the 11-foot wall.
* Clearance: Maintain a 30" x 48" clear floor space on the side of the table for patient transfer.
* The FibroScan: Use a 24" x 24" footprint for the Rolling FibroScan 630. 



The 9x11 Layout Summary 

Feature Measurement
Door Clearance 32" Minimum (36" Preferred) Required for High-Volume Throughput.
IT Drop Height 18" Above Finished Floor (AFF) Ensures Remote PI data cables stay hidden.
Sink Faucet 4" Wrist-Blades or Sensor
Switch Height 15" - 48" AFF Standard ADA Reach Range for COE Certification.
Immediate Action for Your Draftsperson 
Give them this note: 

"The 9' x 11' room must be labeled a 'High-Acuity Diagnostic Suite.' Utilize a wall-mounted floating desk and a pump-assisted wall-hung sink to maintain the 60-inch ADA turning radius. Prioritize three hardwired Cat6 drops in the Hub to support 200-patient data integrity." 
What else to label on the blueprint (The "Wealth" Markers): 
Ask the draftsperson to include these specific sub-labels for each zone to prove "Audit Readiness": 
* Zone A (Center): "High-Acuity Procedure Core (ADA Compliant)"
* Zone B (Left): "Advanced Motility & Diagnostic Bay (Dedicated 20A Power)"
* Zone C (Right): "Surgical Navigation & Remote PI Hub (Hardwired Cat6 Secure Data)"
* Zone D (Entry): "RPM Patient Onboarding Station (200-Patient Capacity)"
* The Sink: "{S} Clinical Sanitation Hub (Pump-Assisted / ADA Compliant)" 


Please place these specific high-acuity assets to prove the room is a "Surgical Feeder" for your ASC. 
Have the draftsperson add these exact items to the "Advanced GI & Hepatology Diagnostics: Center of Excellence" blueprint: 

1. Zone A: The Procedure Core (Center) 
* Asset [ID-001]: Midmark 625 Power Exam Table. Positioned for 360-degree access.
* Asset [ID-002]: Rolling FibroScan 630 Expert.  

2. Zone B: High-Acuity Diagnostic Bay (Left Wall) 
* Asset [ID-003]: Medtronic EndoFLIP™ 300 System.
* Asset [ID-004]: High-Resolution Manometry (HRM) Console.
* Note for Drafter: "Place on (E) Dedicated 20A Medical Power. These consoles identify the $5,000–$8,500 facility fee surgical cases (POEM/TIF) for the ASC." 

3. Zone C: Digital Hub & Sanitation (Right Wall) 
* Asset [ID-005]: Dual 27" 4K Monitors on a wall-mounted floating arm.
* Asset [ID-006]: ADA Wall-Hung Clinical Sink with Saniflo Saniswift pump.
* Note for Drafter: "Install [D] Three Hardwired Cat6 Ethernet Drops centered under the desk at 18" AFF. Mandatory for Remote PI and 200-patient data integrity." 

4. Zone D: CoachCare Hub (Entrance) 
* Asset [ID-007]: CoachCare Cellular Scale and BP Cuff Demo Station.
* Note for Drafter: "Label: RPM Onboarding Station (200-Patient Capacity)." 


Tell the draftsperson to add this specific 2026 Audit Note to the bottom of the plan: 

"This facility is a High-Acuity Diagnostic COE optimized for Surgical Navigation. All diagnostic data is transmitted via hardwired Cat6 data drops to the Remote PI Hub. The room maintains a 60-inch ADA turning radius through the use of wall-mounted fixtures and slim-profile rolling carts."